Key takeaways
- Infrared spectroscopy confirms identity, not purity or concentration.
- Reflectance sampling needs almost no preparation and suits incoming checks.
- A high match score supports identity; it does not prove the correct grade.
- Similar grades of the same material can give nearly identical spectra.
- Pair the spectrum with a physical check such as viscosity or melting range.
An incoming raw material can arrive with the right name on the bag and the wrong substance inside. Infrared spectroscopy answers the identity question in minutes with almost no sample preparation, provided the limits of the technique are understood. It confirms what a material is, not how pure it is.
1. The question it answers
Identity testing asks whether the material in the drum is what the label says. It does not ask whether the assay is 98 or 99 percent, whether water content is within specification, or whether the grade suits the formula. Those need other methods. Confusing the two produces either false confidence or unnecessary testing, and both cost money.
2. How the spectrum is made
Infrared light excites molecular vibrations, and each functional group absorbs at characteristic wavenumbers. The resulting spectrum is a fingerprint of the molecule’s bonds. Two samples of the same substance give the same pattern of peaks, while a different substance gives a different pattern even when both look identical as white powders. That is the value of the method: a name and an appearance can be wrong, a spectrum usually is not.
3. Reference spectra
A result means something only against a reference. The best reference is a retained sample of the approved material, run on the same instrument under the same conditions. Library spectra help with identification but can differ in peak intensity because of sample preparation, so use them to screen and the retained sample to confirm.
4. Sample preparation
Reflectance sampling places the material on a crystal and needs no dilution, which makes it fast and repeatable. Transmission methods need a pressed disc or a solvent film and are more laborious but show finer detail. For incoming checks reflectance is usually the practical choice, with a clean crystal and consistent contact pressure every time. Clean the crystal between samples, because a film left by the previous material shows up as a ghost peak.
5. What it cannot tell you
Peak intensity is not a reliable measure of concentration, because it depends on contact, thickness and instrument state. The method will not detect a small adulterant inside a similar matrix, and it says little about water content, microbial load or heavy metals. A passing spectrum is one gate, not a release decision.
6. Setting a pass threshold
Most software returns a match score against the reference. A threshold near the top of the scale is common, but the number should come from running known-good and known-bad samples and looking at the separation between them. A threshold nobody has challenged is a number, not a control, and auditors ask how it was set. Record the samples used to set it, including the ones expected to fail.
7. Where false passes happen
Similar grades of the same material can differ only in the fingerprint region. Silicone fluids of different viscosities are the classic case: the chemistry is identical and the difference is chain length. Emulsifier blends with the same components in different ratios can also look close. Where grade matters, add a viscosity or melting point check.
8. Pairing with quick tests
Refractive index, melting range, pH of a solution, density and appearance cost minutes and catch many grade errors that infrared misses. A panel of two or three physical checks beside the spectrum gives a far stronger identity confirmation than either alone, and the extra cost is a few minutes per lot. Choose the checks that separate the grades actually bought.
9. Records and audit
Keep the spectrum, the reference used, the operator, the instrument and the date with the batch record. When a complaint arrives two years later, the spectrum is the evidence that the material entering the plant was the approved grade. Store the reference under the same conditions as the raw material, or it will drift and so will the test.
10. Where it fits
Identity checks apply to every raw material behind our creams and essences. Related quality notes continue in encyclopedia.
Reference: Quality control SOP notes, 2025